The Department of Physical Chemistry invites quotation from establised vendors/manufacturer for 1 unit of Digital Storage Oscilloscope with following minimal specification :
TENDER
SPECIFICATIONS FOR 100MHz, 2 CHANNEL, BATTERY/MAINS OPERATED DIGITAL STORAGE
OSCILLOSCOPE WITH COLOR DISPLAY
The Digital Storage
Oscilloscope should have following features:
*2 fully isolated and
floating channels
*8 hours of
continuous battery operation with two batteries
*Optional Power
Application software for Power analysis
*Color LCD display
*Automatic Measurements
*FFT measurements
*External Trigger Input
*Auto set facility
*Waveform and Setup
Memory
*Documentation and
measurement results with Software or integrated Compact Flash Mass storage
SPECIFICATIONS:
SIGNAL
ACQUISITION:
Bandwidth : Single Shot - 100 MHz
Repetitive – 100 MHz
Sample
Rate : 1 GS/s on each Channel
Isolated Channels
: 2
Sensitivity : 2mV to 5V/div
DC Vertical Accuracy : ±3%
Vertical Zoom :
vertically expand or compress a live or stopped waveform
Isolated External Trigger Input :
should be available
Max Input Voltage (1 Mohm) : 300V RMS CAT
II at BNC Signal to BNC Shell, 1000V RMS CAT II At Probe Tip to earth ground
with optional High voltage Probe
Float Voltage :
600V RMS CAT II or 300V RMS CAT III BNC shell to earth ground, 1200V RMS CAT II between any two channel commons with each of
the channels commons not
being more than ±600V RMS from earth groundund
TIME BASE
SYSTEM:
Time/Division
Range : 5nS to 50s/div
Time Base
Accuracy : 50 ppm
Record
Length : 2.5 K points (each channel)
Horizontal
Zoom : Horizontally expand or compress a live or
stopped waveform
NONVOLATILE
STORAGE:
Nonvolatile Storage : Standard (with
Compact Flash Mass Storage)
Reference Waveform
Display : Two 2500 point reference waveforms
Waveform
Storage : 96 or more reference waveforms per 8 MB
Setups : 4000 or more front panel setups
per 8 MB
Screen
Images : 128 or more screen images per 8 MB
Save
All : 12 or more Save all operations per
8 MB
TRIGGERING
SYSTEM:
Trigger
Type : Edge, Video, Pulse Width (or Glitch)
Video Trigger
Type : Trigger on all lines or individual line, odd/even
or all fields, from Composite Video, or
Broadcast standard (NTSC, PAL, SECAM)
Pulse width (or
Glitch) : Trigger on a pulse width less than, greater than,
equal to or not equal to a selectable time
limit ranging from 33 ns to 10 sec.
Trigger
Source : CH1, CH2, Ext, Ext/5 and Ext/10
DISPLAY SYSTEM:
Color : 1/4 VGA, passive color LCD
with color on black background with Adjustable multi-level contrast and brightness
controls
Interpolation : Sin(x)/x
Persistence : Off, 1 sec, 2 sec, 5 sec,
Infinite
I/O interface
Printer Port
: standard
RS-232
port : standard
RS-232
programmability : full talk/listen modes, control of all modes,
setting and measurement
Mass Storage Compact
: to accept any Type 1 Compact Flash card upto 1 GB
Flash memory
Power Measurement
(optional):
Optional package should be available which can offer
instantaneous power waveform analysis, waveform analysis, harmonics
analysis, switching loss, phase angles, dv/dt and di/dt cursors.
GENERAL:
The instrument should
meet all EMC and Safety standards, and to be provided with 2 Nos. 1x/10x
Probes, Operational Manual, Power Cord and NIST Traceable Calibration
Certificate. Optional Current Probes should be available for Power
measurements analysis.
Necessary software
should be provided to connect with PC via RS-232 and data transfer to
Microsoft Word and Microsoft Excel application.
Minimum Warranty for the
instrument should be 5 years and after Sales service facility should
be available.
Authorised Dealership Certificate is must incase of
principal manufacturing company not quoting directly.
|