Sealed tenders are
invited for a Zeta-Potential measurement instruments with the facility of
Particle Size and Molecular Weight Analyzer with suitable Software for
operation through PC, spectra handling etc. and having the following
specifications & Accessories:
Tender Form: Tender
should be submitted to the Registrar, IACS, Jadavpur, Kolkata-32 in plain
paper or Letter Head separately in the forms of Technical and price Bids
both enclosed in a single envelop. Complete Tender Bids in a sealed envelop
should be reached with in 17.12.2010. Payment should be made through
Letter of Credit.
Specifications: The single system should have the measurement facility
for Zeta Potential, Nano particle size and Molecular Weight measurement at
variable temperature (0˚C -90˚C) through Peltier temperature controller.
(a) Zeta Potential
-
Principals: Preferably by high & Low frequency micro
electrophoresis, Laser Doppler Velocimetry with Phase-analysis light
scattering.
-
Size Range suitable for measurement: Min: 5nm, Max: 90
micron
-
Sample Concentration: 40% w/v
-
Zeta Potential range: with no limitation
-
Facility for zeta potential measurement in low as well as
high concentration samples by using Slow Field and Fast Field Reversal.
-
Live measurement windows for display of Frequency,
Voltage and Current.
(b) Particle size
analysis and Molecular weight determination
-
In aqueous and non aqueous solution phase.
-
Minimum Particle size range (dia): Less than 0.5 nm with
sufficient documentary proof.
-
Maximum Particle size range (dia): At least up to 10
micron with sufficient documentary proof
-
Measurement principle: preferably by Non Invasive Back
Scattering.
-
Minimum sample required: less than 15µl.
-
Concentration range: Min 0.1mg/ml 15 kDa
protein --- 40% w/v.
-
Molecular weight range : preferably in the range
of 500 Da to 2 x 107Dalton
Laser : He-Ne Laser
Wave
Length : 633 nm (Visible)
Detector : Avalanche Photo Diode
Detector.
(c) Application Software
- Suitable Application software for Temperature,
Time Trend Analysis
- Intensity, volume, number wise Particle Size
Distribution.
- Protein characterization tool with Protein Melting
point, Protein Crystal Screening,
- 2nd Viral Coefficient determination
- Display result of Zeta Potential, Mobility,
Conductivity, Temp, Iso electric Point.
Any other additional features in favour of your
instrument.
Technical compliance data originally signed should be
included along with quotation.
Essential accessories/items
(i) Computer [2GB RAM, 80GB Hard Disk, 17” monitor
(LCD), Combo drive, Key board, optical mouse, Windows XP OS] and HP LaserJet
printer and UPS (online /line interactive minimum 1KVA with 30 minutes
backup time).
(ii) Standard samples for validation of Zeta Potential,
Particle size measurement.
(iii) Essential Cuvettes for the operation of the above
system.
Installation/Demonstration /Application Training at
site: Free of cost by the supplier
Warranty: Specify time period
User list: The supplier should provide a list of
current users (with contact no, Email address etc) for the quoted item and
if required the bidder should also arrange the demonstration for the quoted
items.
Service facility: Supplier should mention their
details of service setup and Manpower in Kolkata who are responsible for
after sales support. Response time should be within 24 hours
Date of Opening:
Tender will be opened on 27th December,
2010 at 1.00 pm in the JC Bose Meeting
room of IACS
(Updated December 20, 2010)
Acceptance of Tender:
IACS authority does not bind itself to accept the lowest tender and reserves
the right to reject any or all the tenders received without assigning any
reason thereof.
Registrar
|