Sealed
tender in two bids system (Technical Bid &
Financial bid) is invited for Kr Ion Laser
with cooling system and Vibration free Optical Table
having the following specifications & Accessories:
1)
Krypton ion Laser
-
A 4-5 Watt all line
visible krypton ion Laser including Laser head: Must have the following
Lines 406.7 nm, 413.1 nm, 468 nm or 482 nm, 530.9 nm, 568 nm, 647 nm and 676
nm. Additional lines like 415 nm, 488nm and 752 nm may be present.
-
Output power of the 413 nm and 647 nm should
be greater than 1 W and those for the rest of the lines should be greater
than 300 mW immediately after installation of the laser.
-
The Laser tube should
be manufactured by the same company who makes the Laser and not any third
party.
-
Low-Noise Magnet with
a cooling system to minimizes vibration due to water turbulence.
-
Remote controlled external aperture.
-
Power supply.
-
Complete set of prism assembly for all
excitation wavelengths offerred.
-
Complete set of prism assembly for Multiline
operation.
-
Remote control unit with automatic wavelength
search.
-
Proper transformer to operate the Laser.
-
Proper chilling system to operate the Laser.
-
Laser power meter for 10-4-10 Watt
range.
-
The vendor should state
the required utility condition for running the Laser.
2)
Optical Table system
Table
-
Dimension: 1500 mm x 2400 mm x 305 mm
-
Mounting Hole Pattern: 25 mm grid
-
Length: 2400 mm
-
Width: 1500 mm
-
Thickness: 12 in.
-
Ferromagnetic stainless steel working surface
with surface flatness: ±0.1 mm
-
Core Design: Trussed Honeycomb, Vertically
Bonded Closed Cell Construction, Steel sheet materials
-
Hole/Core Sealing: Easy clean conical cup 0.75
in. (19 mm) deep, Non-corrosive high impact polymer material
-
Top and Bottom Skins: 4.8 mm thick with
integrated damping lay
-
Deflection Under Load <1.3 x 10-3 mm
Standard Isolators with
Self Centering capability
-
Height: 711 mm
-
Load per Isolator: 900 kg
-
Number of Isolators: 4
-
Re-leveling Accuracy: ±0.250 mm
-
Vertical Isolation, 5Hz: 94 %
-
Horizontal Isolation, 5Hz: 85 %
-
Max. Air Pressure: 1.4-6.0 kg/cm2
-
Vertical Isolation, Res: 1 Hz
-
Vertical Isolation, 10Hz: 98 %
-
Horizontal Isolation, Res: 1.5 Hz
-
Isolation, Horizontal 10Hz: 95 %
-
Amplification at Resonance, Vertical.: 10 dB
-
Amplification at Resonance, Horizontal: 9 dB
Air compressor with air
intake filter, regulator filter, and automatic turn off switch
-
Operating Sound Level (1ft) (dB) :30
-
Release Value Sound Level (dB) :62
-
Flow Rate (20–80 psi) :0.80 – 0.68
-
Tank Capacity (liters): 3.5
Tender Submission Date: Tender will have to be
submitted in two covers (Technical Bid & Financial bid) separately sealed
and both enclosed in a single sealed envelop. Complete Tender Bids should be
reached to the following address within 05.11.2010.
Opening of Tender: Technical bid will be
opened at 3:30 p.m. in the Council Room of IACS on 08.11.2010
in presence of the bidders or their authorized representatives who may like
to be present. Time and date of opening of financial bid of tender will be
intimated thereafter.
Note:
1.
The Technical Bid should accompany a compliance chart.
2.
Warranty clause, and clauses for on-site operational & routine
maintenance training should be mentioned by the Bidder.
3.
The Bidder will arrange for providing training for using the
equipments at the time of installation to the designated users.
4.
The rate should
be inclusive of all taxes, transportation charges including transit
insurance etc. Nothing extra will be paid in addition to the quoted
rate.
5.
The
Bidder
should mention the model no, make and shall submit a printed literature
of the product with the technical bid positively.
6.
The
Bidder
should mention (with proper documents) the number of installations of the
system in India in last five years.
7.
The
Bidder
should sign on all pages of the tender document along with their company
seal
8.
Guarantee
certificate, users manuals etc. are required to be handed over to the user
after successful commissioning of the system.
9.
The Institute authority does not bind itself to accept the lowest
10.
The Institute reserves the right to reject
any or all of the tenders received without assigning any reason thereof.
11.
The Technical Bid should accompany the
minimum size of the room required and details of site preparation
(electrical, civil etc.) requirements
12.
Conditional/incomplete tenders or tender shall be summarily rejected.
13.
EMD of 2% of the
estimated cost of goods in the form of A/C Payee Demand Draft in the favor
of Indian Association for the Cultivation of Science is required.
14.
Performance
Security of 5% of the estimated cost of goods in the form of A/C Payee DD
in the favor of Indian Association for the Cultivation of Science is
required after qualification of technical specification and will be refunded
after successful installation of the instrument.
Registrar
|