Sealed tenders are invited
from bonafide, resourceful and eligible manufacturers/exclusive
distributors, who are enlisted to Indian Association for the Cultivation of
Science, for a Fluorescence Spectrometer UV-VIS-NIR range
by the Registrar by 14th
February, 2011 upto 5-00 P. M. and will be opened on
16th
February 2011.
17th February, 2011 (Updated
on February
04, 2011) Manufacturers/exclusive distributors should have past
history of supplying this type of instrument to this or other reputed
organizations.
The required specifications
are as follows:
(i) Excitation Source : 450 W xenon short arc lamp or
any other suitable ozone free excitation source. The preferred excitation
wavelength range 230-1000 nm.
(ii) Monochromators: Double grating excitation for
excitation in the wavelength range 230-1000 nm and single/double grating
emission monochoromator in the range upto 1300 nm. Appropriate optical
component to measure up to 1700 nm will be preferred.
(iii) Wavelength Accuracy: :±0.5 nm with sub-nm
resolution and suitable scanning speed.
(iv) Detectors: Excitation correction from 240 nm-1000
nm, high sensitivity in 240nm-850 nm range and to cover upto 1700 nm or
more.
(v) Instrument must include automated Polarizers for
excitation and emission pathways. The polarizers should be fully computer
controlled and can be manually set or automatically controlled for
polarization, anisotropy or G-factor correction is essential.
(vi) The instrument should be compatible with remote
fibre optics measurement. Fibre optic platform with light from excitation
monochromator and redirection to emission port. A microscopic attachment
option will also be preferred.
System should be fully computer controlled with
necessary software for the system control data collection and data analysis.
Software for Nano particle size measurement and analysis would be preferred.
Essential Accessories preferred to be offered with the
main instrument: Solid sample holder For powers and films, Fiber optic
attachment for remote measurement, Filter and Filter holder for second order
cut of.
The tenders should be sent to the following address:
Registrar,
Indian Association for the
Cultivation on Science
2A & 2B, Raja S. C.
Mullick Road, Jadavpur, Kolkata -700032
The tender notice number and the name of the
instrument should clearly be written on the envelope.
Incomplete tenders or tender
shall be summarily rejected.
Registrar
|