(i) |
Mounting: Floor mounted
system for dedicated use in a laboratory. |
|
(ii) |
Sample/Detector Positioning System:
Goniometer: It Should be fully automated
Goniometer. It should have very high angular precision and high
angular coverage of minimum 150 deg. 2 Theta. Colour video
microscope for easy alignment of samples should be an in-built
feature. Video monitor should be inside the enclosure. |
|
(iii) |
X-ray Detector & Optics: (a) X-ray Detector:
Charge Couple Device (CCD) detector
for single crystal frame data collection and imaging/camera,
large active area, very high sensitivity of minimum 160 eV or
higher per Mo photon, having maximum number of pixel for best
resolution and sensitivity, very fast effective read out rate
with very high signal to noise ratio, must be equipped with
adequate cooling facility, (b) Beam Optics: Graphite
Monochromator with all other necessary accessories/optics for
using in both Molybdenum (Mo) and Copper (Cu) radiation.
|
|
|
Powder X-ray data for powder diffraction pattern
recording as well as small angle X-ray scattering data
acquisition:
The system offered should have the possibility of recording
powder X-ray diffraction (PXRD) and small angle X-ray scattering
(SAXS) data from microcrystalline sample filled in a capillary
mounted on the same goniometer of the system. Suitable software
should be provided to convert the frame data collected in the
capillary to typical powder diffraction pattern. Suitable
software should be provided to analyze the SAXS data.
The breakup prices for these facilities should be
shown in the financial bid. |
|
(iv) |
Environment: (a). Operating Temperature:
Approximate: 18 – 30oC, (b). Operating Humidity:
Approximate 5-80% RH. |
|
(v) |
X-ray Generator:
Computer controlled solid state based X-ray generator with power
output of 3KW or above, sealed X-ray generator with maintenance
free remote control operation, high speed / safety shutter and
fail-safe switch, generator stability
± 0.005% per
± 10% variation in line
voltage or better, comprehensive range of alarm and malfunctions
indicator for overloading, tube current, cooling water, line
voltage, and target cooling, adjustable maximum load capability,
display of KV and mA and adjustable through computer.
Maintenance: It must be repairable at our laboratory;
Original manufacturer of the XRD must confirm this in the
quotation. |
|
(vi). |
X-ray Tube: Ceramic
X-ray tubes with Molybdenum (Mo) and Copper (Cu) anode with fine
focus 2000 Watt or more. All necessary optics are to be
included in the basic system so that both the tubes can be used
in the offered system for crystallographic work. While using the
Copper radiation, all necessary optics must be included in the
basic system in order to achieve high sensitivity for the CCD
detector. |
|
(vii) |
Radiation
enclosure: Fully X-ray protected enclosure as per
international safety norms. |
|
(viii) |
Low & high Temperature attachment:
Temperature from 80K – 500K equipped with liquid nitrogen
Dewar of minimum 60 Liters capacity or more, pressure
regulator, transfer line, line heater and necessary accessories;
very low liquid Nitrogen consumption, no icing effect should be
ensured. This facility should be offered with breakup price.
Optionally the vendor should offer the cryosystem with dedicated
LN2 generator. |
|
(ix) |
Computer and Printer-
The computer should be come
from the manufacturer of the SXRD system with factory loaded
software.
Laser Jet Colour Printer. |
|
(x) |
Application software-
WINDOWS XP/ LINUX based application software on GUI platform for
instrumental control for single crystal, twins, low/high
temperature, high pressure, charged density and modulated
structure. The software should be able to perform complete data
acquisition, scaling, space group determination, structure
determination and final report generation for publication
purpose.
No public domain software is acceptable. Manufacturer must offer
their licensed software developed by them. |
|
(xi)
|
Spares: Ceramic fine
focus X-ray tube with molybdenum Anode: One (1) No. In addition,
spares for at least 3 years normal operation to be included in
the offer. The detail list of spare to be enclosed with the
offer for evaluation purpose. Supplier should confirm the
availability of spares for next 10 years from the date of
installation. |
|
(xii) |
Crystal mounting accessories:
Capillaries:
Capillaries made of special glass with wall thickness of 0.01 mm
and outer diameter of 0.2mm, 0.3 mm, 0.5mm – 25 pcs. Of each
type and other necessary capillaries / accessories.
Cryo-loop: 25 cryo-loops
for cryo-mounting. |
|
(xiii) |
Test crystal: One test
crystals to be included in the basic system |
|
(xiv) |
Goniometer Head: 5
(five) Nos. of Goniometer head in the basic system. |
|
(xv) |
Warrantee: The single
crystal X-ray diffractometer system including X-ray tube, water
chiller and UPS quoted for it should be warranted for a period
of 36 months from the date of installation. A 2 year Annual
Maintenance Contract (AMC) after the expirey of warranty (i.e .
after 3 years) should be offered free of charge. |
|
(xvi) |
Qualification criteria:
The data collected in the offered SXRD system
must be publishable as per the Acta Crystallographica guideline.
This is applicable for both Mo and Cu radiation. |
|
(xvii) |
Pre-Installation requirement:
Necessary pre-installation advice should be sent immediately
after the placement of the order. |
|
(xviii) |
Installation, commissioning and Application
Training: Free of cost at site for
minimum 10 working days for a group of technical staff/students
from operating the instrument to complete structure
determination/solution. |
|
(xx) |
Service facility in India :
Supplier should clearly mention about their service set up in
India for prompt service support along with number of service
engineers specially trained on the offered system. |
|
(xxi) |
Installation in India:
Detailed lists of Indian users for the CCD-based XRD quoted
model. There must be at least two same quoted model
installed/ordered in India in last 3 years. |
|
2 |
External water chiller for the single crystal
X-ray diffractometer System:
Both imported and indigenous (ISO certified) chiller for
X-ray generator and tube should be offered with breakup prices.
After sales service should be provided by the Vendor. |
|
3 |
UPS for the single crystal X-ray diffractometer
system: Suitable 3 – phase input, 3 –
phase output on-line UPS with digital signal processor (DSP)
control double conversion with IGBT with latest version and at
least 20 KVA or more rating capacity with sealed maintenance
free batteries, for supporting the XRD system and external water
chiller with minimum 1 hour back up time with detailed technical
specification from reputed brand manufactures. After sales
service should be provided by the Vendor. |
|
4 |
Manuals / Circuit-Diagrams and Instruction
Sheets: All the manuals including
circuit Diagrams and instruction sheets must be supplied in
English for the purpose of service engineer’s reference.
|
|
5 |
Validity of quotation:
Minimum 4 months. |
|
6 |
Tender should be
submitted in two parts – technical and
financial |
|
7 |
Last date of tender
submission: 30th Nov. 2011 |